Proline Prosonic Flow P 500
Endress+Hauser 49
*The distance between the strap bolts and strapping band lock must be at least 500 mm (20 in).
• For 1-traverse mounting with 180° (opposite) → 6, 9 (single-path measurement,
A0044304), → 10, 9(two-path measurement, A0043168)
• For 2-traverse mounting → 7, 9 (single-path measurement, A0044305),
→ 11, 9(two-path measurement, A0043309)
• Electrical connection
Sensor holder with welded bolts )
Can be used for
• Measuring devices with measuring range DN 50 to 4000 (2 to 160")
• Mounting on pipes DN 50 to 4000 (2 to 160")
Procedure:
• The welded bolts must be fastened with the same installation distances as the mounting bolts with
strapping bands. The following sections explain how to align the mounting bolts, depending on
the mounting method and measurement method:
• Installation for measurement via 1 traverse → 51
• Installation for measurement via 2 traverses → 54
• The sensor holder is fastened as standard using a locking nut with a metric M6 ISO thread. If a
different thread is to be used for fastening, a sensor holder with a detachable locking nut must be
used.
A0043375
41 Holder with welded bolts
1 Welding seam
2 Locking nut
3 Hole diameter max. 8.7 mm (0.34 in)
Installing sensor – small nominal diameters DN 15 to 65 (½ to 2½")
Requirements
• The installation distance is known. → 44
• Sensor holder is pre-assembled.
Material
The following material is required for mounting:
• Sensor incl. adapter cable
• Sensor cable for connection to the transmitter
• Coupling medium (coupling pad or coupling gel) for an acoustic connection between the sensor
and pipe