Proline t-mass 65 PROFIBUS DP/PA Technical data
Endress+Hauser 109
10.1.5 Power supply
Electrical connections ä 27
Supply voltage 85 to 260 V AC, 45 to 65 Hz
20 to 55 V AC, 45 to 65 Hz
16 to 62 V DC
Cable entry Power supply and signal cables (inputs/outputs):
Cable entry M20 × 1.5 (8 to 12 mm (0.31 to 0.47 inch))
Threads for cable entries, ½" NPT, G ½"
Connecting cable for remote version:
Cable entry M20 × 1.5 (8 to 12 mm (0.31 to 0.47 inch))
Threads for cable entries, ½" NPT, G ½"
Cable specifications
(remote version)
ä 29
Power consumption AC: 85 to 260 V = 18.2 W ; 20 to 55 V = 14 W ; (including sensor)
DC: 8 W (including sensor)
Power supply failure Lasting min. 1 power cycle:
EEPROM/HistoROM T-DAT saves measuring system data if power supply fails.
HistoROM S-DAT is an exchangeable data storage chip with sensor specific data: (pipe type,
nominal diameter, serial number, flow conditioner, zero point, etc).
Totalizer stops at the last value determined
Potential equalisation No measures necessary.
For devices in hazardous areas please refer to the additional Ex documentation.
10.1.6 Performance characteristics
Calibration reference
conditions
Accredited according to ISO/IEC 17025
Traceable to National Standards
Calibration gas: air
Temperature controlled to 24 °C ±0.5 °C (75.2 °F ± 0.9 °F) at atmospheric pressure
Humidity controlled < 40% RH
Maximum measured error t-mass 65F:
±1.5 % of reading for 100 % to 10 % of full scale (at reference conditions)
±0.15 % of full scale for 10 % to 1 % of full scale (at reference conditions)
t-mass 65I:
±1.0% of reading
±0.5% of full scale (at reference conditions)
Repeatability ±0.4 % of reading for velocities above 1.0 m/s (3.3 ft/s)
Influence of medium pressure
(Pressure co-efficient)
Air: 0.35% per bar (0.02% per psi) of process pressure change
Other gases: on request
Response time Typically less than 2 seconds for 63 % of a given step change (in either direction).