Trouble-shooting Proline t-mass 65 PROFIBUS DP/PA
92 Endress+Hauser
9.2.2 List of system error messages
No. Device status message
(local display)
PROFIBUS measured value status
Extended
diagnostic
message in the
PROFIBUS
master
Remedy / spare parts
Spare parts ä 99
Quality code (HEX)
Measured variable status
Quality status
Quality substatus
Limits
S = System error
$ = Fault message (with an effect on the current operation)
! = Notice message (without any effect on the current operation)
No. # 0xx Hardware error
001 S: CRITICAL FAIL
$: # 001
0x0F BAD Device Failure Constant ROM/RAM
failure
Cause:
Serious device error
Remedy:
Replace the amplifier board.
011 S: AMP HW EEPROM
$: # 011
0x0F BAD Device Failure Constant Amplifier EEPROM
failure
Cause:
Amplifier: Defective EEPROM
Remedy:
Replace the amplifier board.
012 S: AMP SW EEPROM
$: # 012
0x0F BAD Device Failure Constant Amp. EEPROM
data inconsistent
Cause:
Measuring amplifier: Error when accessing data of
the EEPROM
Remedy:
Replace the amplifier board.
014 S: AMP SW-ROM/RAM
$: # 014
0x0F BAD Device Failure Constant Amp. ROM/RAM
data inconsistent
Cause:
Amplifier: Defective ROM/RAM
Remedy:
Replace the amplifier board.
031 S: SENSOR HW DAT
$: # 031
0x10
0x11
0x12
BAD Sensor Failure O.K.
Low
High
S-DAT failure / not
inserted
Cause:
Sensor DAT:
1. HistoROM/S-DAT is defective.
2. HistoROM/S-DAT is not plugged into the
amplifier board or is missing.
Remedy:
1. Replace the S-DAT.
Check the spare parts set number to ensure
that the new, replacement DAT is compatible
with the measuring electronics.
2. Plug the S-DAT into the amplifier board
ä 100, ä 102
032 S: SENSOR SW DAT
$: # 032
0x10
0x11
0x12
BAD Sensor Failure O.K.
Low
High
S-DAT data
inconsistent
Cause:
Error accessing the calibration values stored in the
HistoROM/S-DAT.
Remedy:
1. Check whether the HistoROM/S-DAT is
correctly plugged into the amplifier
board ä 100, ä 102
2. Replace the S-DAT if it is defective.
Before replacing the DAT, check that the new,
replacement DAT is compatible with the
measuring electronics. Check the:
– Spare parts set number
– Hardware revision code
3. Replace measuring electronics boards if
necessary.