Technical data Proline t-mass 65 PROFIBUS DP/PA
110 Endress+Hauser
10.1.7 Operating conditions: Installation
Installation instructions ä 12
Inlet and outlet runs ä 15
Length of connecting cable Max. 100 m (328 feet), remote version
System pressure ä 12
10.1.8 Operating conditions: Environment
Ambient temperature Standard: –20 to +60 °C (–4 to +140 °F)
Optional: –40 to +60 °C (-40 to +140°F)
!
Note!
• Install the device in a shady location. Avoid direct sunlight, particularly in warm climatic regions.
(A protective sun cover is available on request)
• At ambient temperatures below –20 °C (–4 °F) the readability of the display may be impaired.
Storage temperature –40 to +80 °C (–40 to +176 °F), recommended +20 °C (+68 °F)
Degree of protection Standard: IP 67 (NEMA 4X) for transmitter and sensor
Shock resistance According to IEC 60068-2-31
Vibration resistance Acceleration up to 1 g, 10 to 150 Hz, following IEC 60068-2-6
Electromagnetic compatibility
(EMC)
To IEC/EN 61326 and NAMUR recommendation NE 21