Installation Smartec S CLD132
16 Endress+Hauser
C07-CLD132xx-11-06-00-en-008.eps
Fig. 13: Post mounting of CLD132 separate version
3.4.2 Mounting CLD132 compact version or CLS52 sensor for
separate version
Install the compact version or the CLS52 sensor directly on the pipe or vessel socket via the process
connection (depending on ordered version).
!
Note!
Perform an Airset and calibrate the sensor before installing the compact version or the sensor.
C07-CLD132-11-06-00-en-006.eps
Fig. 14: Installation of CLD132 compact version
1. When installing the Smartec S CLD132 or the sensor, make sure that the flow opening of the
sensor is oriented in the flow direction of the medium. An orientation arrow on the sensor
facilitates orientation (see Fig. 14 above).
2. Tighten the flange.
3. For versions with internal thread G 1½, expansion bellows are supplied for length
compensation. Thus, the sensor can always be oriented in flow direction.
!
Note!
Choose the immersion depth of the sensor in the medium such that the coil body is completely
immersed.
Please observe the notes on the wall distance in the chapter "Installation conditions".
Please observe the limits for the medium and ambient temperature when using the compact
version (see chapter "Technical data").
max. 60 / 2.36
mm / inch
min. DN 40*
min. DN 40*
Flow direction
Flow
direction
Orientation
arrow
Horizontal flow Vertical flow