Smartec S CLD132 Commissioning
Endress+Hauser 45
6.4.9 Service
K7
Enter uncompensated
conductivity
0.0 mS/cm
0.0 ... 9999 mS/cm
K8
Enter concentration
value for K6
0.00 %
0.00 ... 99.99 %
K9
Enter temperature
value for K6
0.0 °C
-35.0 ... 250.0 °C
K10
Enter whether or not
the table status is ok
yes
no
Back to K.
Coding Field Selection or range
(factory settings bold)
Display Info
K7
mS/cm
conduct.
0.0
K8
%
concentr.
0.0
K9
°C
Temp.val.
0.0
K10
Status ok
yes
Coding Field Selection or range
(factory settings bold)
Display Info
S
Function group
SERVICE
Settings for service functions.
S1 Select language
ENG = English
GER = German
FRA = French
ITA = Italian
NEL = Dutch
ESP = Spanish
This field must be configured once during
start-up. Then you can exit S1 and continue.
S2 HOLD effect
froz. = last value
fix = fixed value
froz.: Display of last value before activation of
hold.
fix: When hold is active, the fixed value
entered in S3 is displayed.
S3 Enter fixed value
0
0 ... 100 %
(of current output value)
Only available if S2 = fixed value.
S4 Hold configuration
S+C = setup and calibration
CAL = calibration
Setup = setup
none = no hold
S = setup
C = calibration
S
SERVICE
S1
Language
ENG
S2
Holdeffec
froz.
S3
%
Fixed Val
0
S4
Auto HOLD
S+C