Smartec S CLD132 Commissioning
Endress+Hauser 53
6.5 Communication interfaces
Please refer to separate operating instructions BA 212C/07/en (HART) or BA 213C/07/en
(PROFIBUS) for transmitters equipped with a communication interface.
C1 (3)
Calibration with
sensor adaptation for
inductive sensors
Airs = Airset (1)
Cellc = cell constant (2)
InstF = installation
factor (3)
Sensor calibration with compensation of wall
influence.
The distance from the sensor to the pipe wall
and the pipe material (conductive or
nonconductive) influence the measured
value. The installation factor compensates
this influence. See chapter "Installation
conditions".
The sensor is installed in the process.
C131
Enter process
temperature (MTC)
25 °C
-35.0 ... 250.0 °C
Only exists if B1 = fixed.
C132
Enter
α value of the
calibration solution
2.10 %/K
0.00 ... 20.00 %/K
This value is specified in the Technical
Information of all E+H calibration solutions.
You can also use the printed-on table to
calculate the value.
Set
α to 0 for calibration with
uncompensated values.
C133
Enter correct
conductivity value of
the medium
current measured value
0.0 ... 9999 mS/cm
Determine the correct conductivity value by
a reference measurement.
C134
Calculated installation
factor is displayed
1
0.10 ... 5.00
The distance of the sensor to the pipe wall
and the pipe material (conductive or
nonconductive) influence the measured
value. The installation factor compensates
this influence. See chapter "Installation
conditions".
C135
Calibration status is
displayed
o.k.
E xxx
If the calibration status is not o.k., the second
display line shows an explanation of the
error.
C136
Store calibration
results?
yes
no
new
If C135 = E xxx, then only no or new.
If new, return to C.
If yes/no, return to “Measurement”.
Coding Field Selection or range
(factory settings bold)
Display Info
C1
Calibrat
InstF
~
~
~
~
~
C131
°C
MTC temp.
25.0
C132
%/K
alpha val
2.10
C133
mS/cm
Real val.
10.30
C134
InstFact
1
C135
Status
o.k.
C136
Store
yes