Operation Smartec S CLD132
24 Endress+Hauser
5.2.2 Operating elements
The operating keys are located underneath the housing cover. The display and the alarm LED are
visible through the viewing window. For operation, open the housing cover by removing the 4
screws.
C07-CLD132xx-19-06-00-xx-001.eps
Fig. 25: Operating elements of Smartec S CLD132
1 Liquid crystal display showing measured values and configuration data
2 4 operating keys for calibration and instrument configuration
3 Field for user labeling
4 LED indicator for alarm function
5.2.3 Key assignment
ENDRESS+HAUSER
SMARTEC S
ALARM
CAL
E
+
–
2000
mS/cm
O213
20 mA
1
2
3
4
CAL key
When the CAL key is pressed, the instrument prompts for the calibration
access code:
• Code 22 for calibration
• Code 0 or any other number to view the calibration data
Use the CAL key to acknowledge calibration data and to continue through
the calibration process.
ENTER key
When the ENTER key is pressed, the instrument prompts for the setup
access code:
• Code 22 for setup and configuration
• Code 0 or any other number to view the configuration data.
The ENTER key has several functions:
• It calls up the setup menus from the measuring mode
• It is used to store (acknowledge) data entered in setup mode
• It is used to move on within function groups