Smartec S CLD132 Installation
Endress+Hauser 17
Sensor positioning: compact version
The sensor in the compact housing must be oriented in the flow direction.
If you need to reorient the sensor in relation to the housing, proceed as follows:
1. Remove the cover.
2. Loosen the screws of the electronics box and carefully remove the box from the housing.
3. Loosen the three sensor fastening screws until the sensor can be turned.
4. Align the sensor and tighten the screws. Do not exceed the maximum torque of 1.5 Nm!
5. Reassemble the transmitter housing in reverse sequence of operations.
!
Note!
For exact positions of the electronics box and the sensor screws, see the exploded view in the
chapter "Spare parts".
C07-CLD132xx-11-06-05-xx-010.eps
Fig. 15: Sensor orientation in the transmitter housing
AStandard orientation
B Sensor turned by 90°
3.5 Post-installation check
• After installation, check the measuring system for damages.
• Check the sensor orientation to the flow direction of the medium.
• Check that the coil body of the sensor is completely immersed in the medium.
A
B