6.4.7 Temperature compensation with table
This function group is used to perform a temperature compensation with table (field B2 in SETUP 2
function group).
Enter the
α−T value pairs in the fields T5 and T6.
R5 Enter dropout delay
0 s
0 ... 2000 s
R6 Select simulation
auto
manual
This selection can only be made if limit has
been selected in R1.
R7 Switch relay on or off
on
off
This selection can only be made if manual has
been selected in R6. The relay can be
switched on and off.
Coding Field Selection or range
(factory settings bold)
Display Info
s
R5
Off Delay
0
R6
Simulat.
auto
R7
Relay
off
Coding Field Selection or range
(factory settings bold)
Display Info
T
Function group
ALPHA TABLE
Settings for temperature compensation.
T1 Select table
1
1 ... 4
Selection of table to be edited.
Options 1 ... 4 are only available if the
instrument is equipped with the remote
measuring range switching.
T2 Select table option
read
edit
T3
Enter number of table
value pairs
1
1 ... 10
Up to 10 value pairs can be entered in the
α
table. These are numbered from 1 ... 10 and
can be edited individually or in sequence.
T4 Select table value pair
1
1 ... number of table value pairs
assign
If "assign", go to T8.
T
ALPHA TAB
T1
editCurve
1
T2
Sel.Table
read
T3
No. Elem.
1
T4
Sel.Elem.
1