Smartec S CLD132 Commissioning
Endress+Hauser 37
6.4.4 Alarm
The ALARM function group is used to define various alarms and to set output contacts.
Each individual error can be defined to be effective or not (at the contact or as an error current).
O213
20 mA value:
enter corresponding
measured value
Cond: 2000 mS/cm
Conc: 99.99 %
Temp.: 60.0 °C
entire measuring range
Enter the measured value corresponding to
the maximum current value (20 mA) at the
transmitter output.
Display format from A3.
(Spreading: see Technical data.)
O2 (2)
Current output
simulation
lin = linear (1)
sim = simulation (2)
The simulation is terminated by selecting (1).
O221 Enter simulation value
current value
0.00 ... 22.00 mA
The current value entered here is output
through the current output.
Coding Field Selection or range
(factory settings bold)
Display Info
O213
mS/cm
20 mA
2000
O2
Sel.Type
sim
O221
mA
Simulat.
4.00
Coding Field Selection or range
(factory settings bold)
Display Info
F
Function group
ALARM
Alarm function settings.
F1 Select contact type
Stead = steady contact
Fleet = fleeting contact
The contact type selected here only applies to
the alarm contact.
F2 Select time unit
s
min
F3 Enter alarm delay
0 s (min)
0 ... 2000 s (min)
Depending on the unit selected in F2, the
alarm delay is entered in s or min.
The alarm delay does not affect the LED; it
indicates the alarm immediately
F4 Select error current
22 mA
2.4 mA
This selection must be made even if all error
messages are suppressed in F5.
"
Caution!
If you selected the "0-20 mA" range in O211,
you may not select the "2.4 mA" option here.
F
ALARM
F1
Cont.Type
Stead
F2
Time unit
s
F3
s
Err.Delay
0
F4
Err.Curr
22mA