Commissioning Smartec S CLD132
50 Endress+Hauser
!
Note!
If remote parameter set switching is selected, the parameter sets that have been entered are
processed internally but the fields A1, B1, B3, R2, K1, O212, O213 show the values of the first
measuring range.
M4 Select operating mode
cond = conductivity
conc = concentration
The operating mode can be individually
defined for each parameter set.
M5 Select medium
NaOH, H2SO4, H3PO4, HNO3
Tab 1 ... 4
Only available if M4 = conc.
M6
Select temperature
compensation
none, lin, NaCl,
Tab 1 ... 4 if M4 = cond
Only available if M4 = cond.
M7 Enter
α value
2.10 %/K
0 ... 20 %/K
Can only be entered if M6 = lin.
M8
Enter measured value
for 0/4 mA value
Cond.: 0 ... 2000 mS/cm
Conc.: Unit: A2, format: A3
M9
Enter measured value
for 20 mA value
Cond.: 0 ... 2000 mS/cm
Conc.: Unit: A2, format: A3
M10
Enter switch-on point
for limit
Cond.: 0 ... 2000 mS/cm
Conc.: Unit: A2, format: A3
M11
Enter switch-off point
for limit
Cond.: 0 ... 2000 mS/cm
Conc.: Unit: A2, format: A3
The switch-off point entry selects a max
contact (switch-off point < switch-on point) or
a min contact (switch-off point > switch-on
point), thereby implementing an always
required hysteresis function. Never set the
switch-off point and the switch-on point to
the same value.
Coding Field Selection or range
(factory settings bold)
Display Info
M4
Oper.Mode
cond.
M5
Conc.Tab.
NaOH
M6
TempComp
lin
M7
%/K
alpha val
2.10
M8
µS/cm
0/4 mA
0
M9
mS/cm
20 mA
2000
M10
mS/cm
PV on
2000
M11
mS/cm
PV off
2000