Installation Smartec S CLD134
12 Endress+Hauser
3.2 Incoming acceptance, transport, storage
Make sure the packaging is undamaged!
Inform the supplier about damage to the packaging. Keep the damaged packaging until the matter
has been settled.
Make sure the contents are undamaged!
Inform the supplier about damage to the delivery contents. Keep the damaged products until the
matter has been settled.
Check that the scope of delivery is complete and agrees with your order and the shipping
documents.
The packaging material used to store or to transport the product must provide shock protection
and humidity protection. The original packaging offers the best protection. Also, keep to the
approved ambient conditions (see "Technical data").
If you have any questions, please contact your supplier or your sales center responsible.
3.3 Installation conditions
3.3.1 Notes on installation
Installation positions
The sensor has to be immersed completely into the media. Avoid bubbles in the area of the sensor.
!
Note!
For use in hygienic applications only use materials that comply with 3-A standards 74-03 and the
FDA requirements. The cleanability of a sensor also depends on the way of installation. To install
the sensor in a pipe system use the appropriate and EHEDG certified flow assembly for the
respective process connection.
a0005442
Fig. 3: Installation positions of conductivity sensors
A Not for hygienic applications
Air set
Perform an air set before sensor installation (see chapter "Calibration"). Make sure that the
instrument is ready for operation, i.e. mains and sensor are connected.
A