Appendix Smartec S CLD134
80 Endress+Hauser
11 Appendix
Operating matrix
a0005699-en
C1 (3)
C1 (2)
C1 (1)
C112
C113
C111
C132
C131
A
A1
A4
A3
A2
B
B1
B3
B4
C121
F
F1
F2
F3
F4
F5
O
O1
O211
O212
O213
P
C114
C134
C124
A5
B5
C
Function group
CALIBRATION
Calibration
InstF
= installation factor
Cellc
= cell constant
Airs = Airset
Entry of
calibration temperature
(if B1 = fixed)
–35.0 ... +250.0 C
25.0 C°
°
Entry of a value of
calibration solution
0.00 ... 20.00 %/K
2.10 %/K
Entry of correct
conductivity value of
calibration solution
0.0 µS/cm ... 9999 mS/cm
current meas. value
C133
Entry of
calibration temperature
(if B1 = fixed)
–10.0 ... +150.0 C
25.0 C
.
°
°
Entry of a value of
calibration solution
0.00 ... 20.00 %/K
2.10 %/K
.
C122
Entry of correct
conductivity value of
calibration solution
0.0 mS/cm ... 9999 mS/cm
current meas. value
.
C123
Residual coupling
Start calibration
current meas. value
Display of
residual coupling value
-80.0 ... 80.0 µS
Display of
calibration status
o.k.;
E---
.
Display of calculated
installation factor
0.10 ... 5.0
1.0
Display of calculated
cell constant
0.1 ... 9.99 cm
-1
Store
calibration results
; no; newyes
Edit mode:
Code 22
Read mode:
any code
MEAS. VALUE DISPLAY
Conductivity
and
temperature (°C)
Display of
conductivity and
temperature (°F)
Display of
conductivity
Display of
conductivity (uncompensated)
concentration
Display of current
parameter set
(Remote parameter set
switching only)
Error display
(up to 10 errors)
Err ---
Edit mode:
Code 22
Read mode:
any code
Function group
SETUP 1
Selection of operating mode
conc = concentration
cond = conductivity
Selection of display
unit
; mg/l; %; TDS; noneppm
Selection of display format
(if A1 = conc)
X.xxx; ; XXX.x; XXXXXX.xx
Selection of display
unit
; µS/cm; mS/cm; S/cm;
µS/m; mS/m; S/m
auto
Function group
SETUP 2
Selection of
temperature measurement
Pt100
Pt1k (= Pt 1000)
NTC30 (= NTC 30 kW
)
fixed
Selection of temperature
compensation type
none
NaCI = common salt
Tab = table 1 ... 4
(>1 with software option
only)
lin = linear
.
B2
Entry of a value
(if B2 = linear)
0.00 ... 20.00 %/K
2.10 %/K
Entry of correct
process temperature
(if B1 = fixed)
25.0 °C
–35.0 °C ... +250.0 °C
Entry of cell constant
0.1 ... ... 99.99 cm6.3
-1
Temperature sensor
offset
(not if B1 = fixed)
–35.0 ... +250.0
Entry of actual temp.
°C
Function group
OUTPUT
Selection of
current output
; Out 2Out 1
Selection of characteristic
sim = simulation
lin = linear
O2 (2)
O2 (1)
Entry of
simulation value
0 ... 22.00 mA
current value
O221
Selection of
current range
;0-20mA4-20 mA
Entry of
0/4 mA value
entire meas. range
0 µS/cm; 0 %; 0 °C
Entry of
20 mA value
entire meas. range
2000 mS/cm; 99.99 %;
150.0 °C
Function group
ALARM
Selection of
contact type
= steady contact
Fleet = fleeting contact
Stead
Selection of
unit for
alarm delay
;mins
Entry of
alarm delay
... 2000 s (min)
(depending on F2)
0s
Determination of
error current
2.4 mA
22 mA
Selection of
error number
1 ... 255
1
Function group
CHECK
(with software option
only)
PCS alarm setting
(live check)
/1h/2h/4hoff
P1
Monitoring limit
0.3 % of mean value
over time entered