Smartec S CLD134 Technical Data
Endress+Hauser 79
10.9 Chemical durability of CLS54 sensor
No responsibility is taken for the correctness of this information.
Permissible temperature ranges of Smartec S CLD134
a0005499-en
Fig. 43: Permissible temperature ranges of Smartec S CLD134
A CLS54 sensor with separate version
B Compact version
C Short term for sterilisation (<60 min)
Pressure-temperature load curve of CLS54 sensor
a0004809
Fig. 44: Permissible pressure and temperature ranges of CLS54 sensor
A Short-term for sterilisation (< 60 min)
Medium Concentration PEEK
Caustic soda NaOH 0 to 15 % 20 to 90 °C (68 to 194 °F)
Nitric acid HNO
3
0 to 25 % 20 to 90 °C (68 to 194 °F)
Phosphoric acid H
3
PO
4
0 to 15 % 20 to 80 °C (68 to 176 °F)
Sulphuric acid H
2
SO
4
0 to 30 % 20 °C (68 °F)
Peracetic acid H
3
C-CO-OOH 0.2 % 20 °C (68 °F)
60
80
100
120
140
20
400
10
0
–10
–10
20
30
40
50
55
55
35
60
C]
C]
A
B
C
70
Medium temperature
Ambient temperature
95
131
131
257
284
[°F]
[°F]
32
68
104 140 176
212
248
32
50
68
86
104
122
140
158
A
50 70
110 150
10 30
125
90
0
–10
C]
4
8
12
[bar]
32
50
86
122
158
230
302194
257
[°F]
58
116
174
[psi]

Table of Contents