Commissioning Smartec S CLD134
50 Endress+Hauser
6.4.9 Service
K7
Enter uncompensated
conductivity
0.0 mS/cm
0.0 ... 9999 mS/cm
K8
Enter concentration
value for K6
0.00 %
0.00 ... 99.99 %
K9
Enter temperature
value for K6
0.0 °C
-35.0 ... 250.0 °C
K10
Enter whether or not
the table status is ok
yes
no
Back to K.
Coding Field Selection or range
(factory settings bold)
Display Info
Coding Field Selection or range
(factory settings bold)
Display Info
S
Function group
SERVICE
Settings for service functions.
S1 Select language
ENG = English
GER = German
FRA = French
ITA = Italian
NEL = Dutch
ESP = Spanish
This field must be configured once during
start-up. Then you can exit S1 and continue.
S2 HOLD effect
froz. = last value
fix = fixed value
froz.: Display of last value before activation of
hold.
fix: When hold is active, the fixed value
entered in S3 is displayed.
S3 Enter fixed value
0
0 ... 100 %
(of current output value)
Only available if S2 = fixed value.
S4 Hold configuration
S+C = setup and calibration
CAL = calibration
Setup = setup
none = no hold
S = setup
C = calibration