Smartec S CLD134 Commissioning
Endress+Hauser 43
6.4.5 Check
PCS alarm (Process Check System)
The PCS alarm is only available for transmitters with remote parameter set switching.
This function is used to examine the measuring signal for deviations. If the measuring signal is
constant for a specific period of time (several measured values), an alarm is issued. This type of
sensor behaviour may be caused by soiling, etc.
C07-CLD132xx-05-06-00-xx-007.eps
Fig. 37: PCS alarm (live check)
A Constant measuring signal = alarm is triggered after the configured PCS period
!
Note!
An active PCS alarm is automatically cleared when the measuring signal changes.
F5 Select error
1
1 ... 255
Select the errors that are to trigger an alarm
signal.
The errors are selected via the error number.
Please refer to the table in chapter 9.2 "System
error messages" for the error numbers. The
factory settings remain in effect for all errors
not edited.
F6
Set alarm contact to be
effective for selected
error
yes
no
If set to “no”, all the other alarm settings (e.g.
alarm delay) are also deactivated. The settings
themselves are retained.
This setting only applies to the error selected
in F5.
Factory setting is no starting with E080!
F7
Set error current to be
effective for selected
error
no
yes
The error current selected in F4 becomes
effective or is suppressed when an error
occurs.
This setting only applies to the error selected
in F5.
F8
Return to menu or
select next error
next = next erorr
R
If next is selected, the software returns to F5.
If R is selected, it returns to F.
Coding Field Selection or range
(factory settings bold)
Display Info
F5
Sel.Error
1
yes F6
Rel.Assg
F7
Curr.Assg
no
F8
Select
R
A
t