Smartec S CLD134 Commissioning
Endress+Hauser 41
6.4.3 Current outputs
The OUTPUT function group is used to configure the individual outputs.
Furthermore, a current output value can be simulated to check the current outputs (O2 (2)).
B4
Enter process
temperature
25 °C
-10.0 ... 150.0 °C
Only if B1 = fixed.
This value can only be specified in °C.
B5
Display temperature
and calibrate
temperature sensor
Display and entry of real
temperature
-35.0 ... 250.0 °C
This entry is used to calibrate the temperature
sensor to an external measurement.
Omitted if B1 = fixed.
B6
Temperature
difference is displayed
0.0 °C
-5.0 ... 5.0 °C
The difference between the entered actual
value and the measured temperature is
displayed.
Omitted if B1 = fixed.
Coding Field Selection or range
(factory settings bold)
Display Info
B4
C
°
ProcTemp.
25.0
B5
C
°
RealTemp.
0.0
B6
C
°
TempOffs.
0.0
Coding Field Selection or range
(factory settings bold)
Display Info
O
Function group
OUTPUT
Configuration of the current output (not
available for PROFIBUS versions).
O1 Select current output
out1
out2
A different characteristic can be selected for
each output.
O2 O2 (1)
Enter linear
characteristic
lin = linear (1)
sim = simulation (2)
The slope of the characteristic may be
positive or negative.
O211 Select current range
4 ... 20 mA
0 ... 20 mA
O212
0/4 mA value:
enter corresponding
measured value
Cond: 0.00 μS/cm
Conc: 0.00 %
Temp.: -10.0 °C
entire measuring range
Enter the measured value corresponding to
the minimum current value (0/4 mA) at the
transmitter output.
Display format from A3.
(Spreading: see Technical data.)
O
OUTPUT
O1
Sel. Out
out1
O2
Sel.Type
lin
O211
Sel.Range
4-20
O212
µS/cm
0/4mA
0