Commissioning Smartec S CLD134
34 Endress+Hauser
Factory settings
When the instrument is switched on for the first time, the factory settings are in effect. The
following table provides an overview of all major settings.
Please refer to the description of the individual functions in the chapter "Instrument configuration"
for all other factory settings (the factory settings are printed in bold letters).
* if equipped accordingly
Function Factory setting
Type of measurement Inductive conductivity measurement,
temperature measurement in °C
Temperature compensation type Linear with reference temperature 25 °C / 77 °F
Temperature compensation Automatic (ATC on)
Relay function Alarm
Hold Active during configuration and calibration
Measuring range 10 μS/cm ... 2000 mS/cm (measuring range set automatically)
Current outputs 1* and 2* 4 ... 20 mA
Current output 1: measured value for 4 mA
signal current*
0 μS/cm
Current output 1: measured value for 20 mA
signal current*
2000 mS/cm
Current output 2: measured value for 4 mA
signal current*
0.0 °C / 32 °F
Current output 2: measured value for 20 mA
signal current*
150.0 °C / 302 °F