NS 5.14 Firmware
3–22 4900002262 rev. B 3-23-21
• Common Weight: Ratio of the measured spectrum that did not
match a reference spectrum.
• Concentration: Current measured (live) analyte concentration
(process or validation) in selected engineering units.
• Concentration (ppmv): Current measured (live) analyte
concentration (process or validation) in parts per million by volume
(ppmv).
• Concentration Process: Measured analyte concentration of last
process reading in selected engineering units.
• Concentration Unit: Designates the display units for the measured
concentration.
• Conc Process ppmv: Measured analyte concentration of last
process reading in parts per million by volume (ppmv).
• Conversion Type: Sets the type of relations (ideal or real gas) used
in the dew point calculation [0 for ideal gas relations or 1 for real gas
relations (
Z calculated using Peng-Robinson EOS)].
• Cross Shift: The amount shift applied to match the measured
spectrum to the reference spectrum when using Cross Correlation.
• Current 2F Flag: Displays the current scrubber protection usage
level: 0 = standard protection (used for non-differential units), 1 =
mid protection, 2 = high protection, 3 = maximum protection (no
usage).
• Current Midpoint: Current midpoint that the analyzer is using,
including any peak tracking adjustments.
• Custom Precision: Sets the number of viewable digits to the right
of the decimal point.
• Daily Validation: Enables or disables the time of day autovalidation
feature.
• DC Level: Signal level at the high end of the current ramp.
• Dew Point: Measured analyte concentration of last process reading
as a moisture dew point.
• Dew Point Method: Sets the industry standard method with which
the dew point temperature is determined when Calculate Dew
Point is enabled.
• DO Alarm Setup: Bitwise value that sets the functionality of the
Assignable Alarm, as listed in Table 2–3 on page 2–22.
• Dry DC Level: Signal level at the high end of the current ramp from
the last scrubber cycle.
• Dry Pressure (mb): Measured pressure (0.0 for non-differential
units) of the gas sample from the last scrubber cycle in millibar (mb).