ERS
®
Manual / AirCool
®
plus
-System / SP72_TA1C8 / -55°C +200°C / Ver. 02.02 12
3.4 System restrictions
ÿ At intial operation of the whole system (chucksystem and probe station) this whole
system has to be purged with dry air for a minimum time period of 24 hours.
Therefor please close all covers of the probeshield and let the AirCool
®
plus
Chuck System operate at +25°C set temperature.
This procedure is evident of removing any huminidy inside the probe station and
should be repeated after any longer period of down time.
ÿ The dew point inside the probe station will degrade in case a testhead is
provided with fans and/or the probecard area is exposed to air draft (e.g. air-
condition system).
ÿ The dew point sensor has to be recalibrated yearly.
3.5 Temperature-Time diagram
Fig. 3.5a Heating/Cooling curve