EasyManua.ls Logo

ETAS ES930.1 - Page 94

ETAS ES930.1
139 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ETAS Technical Data
ES930.1 - User Guide 94
13.9.4 Digital Input Channels ("DI" Connection)
Number of channels 4
Input voltage 0 V to 5 V
Switching thresholds TTL-compatible,
(high: 2.2 V typ., low: 1.2 V typ.)
Sampling rate 0.5 samples/s to 10 ksamples/s, addi-
tionally events; configurable per chan-
nel
Time resolution (clock rate) 15 ns (66.667 MHz)
Counter 32 bit
Pulse width 120
ns to 64.4
s (= [2
32
- 1] * 15
ns);
resolution 15
ns
Frequency range 0.02 Hz to 3 MHz
Glitch filter Digital software filter;
configurable per channel:
120 ns to 3000 ns or "Off"; resolution
15
ns
Timeout Configurable per channel:
0.1 s to 64.4 s or "Off"
Input impedance typ. 1.5 M || 50 pF
Maximum input voltage Input:
±50 V DC
Input-to-ground of voltage supply:
60 V DC / 30 V AC
Input-to-housing:
60 V DC / 30 V AC
Electrical isolation All 4 inputs together as channel group
to supply voltage and to housing
Measuring functions Event count, active time, inactive time,
status, measurement across several
cycles; parallel per channel; with over
-
flow or timeout control
Events 4 event sources per module can be
configured differently
Events for triggering a synchronous
data acquisition on the ES930.1 mod
-
ule
Events for triggering the model trigger
in the ES910 prototyping module

Table of Contents