E8_Aus_rev07 2021-02-02
48
Do NOT use disinfecting substances to clean the chamber.
Seal and door
Clean the seal and the door with a damp cloth (Fig. 9.2-3), dampened with water or vinegar, to eliminate
traces of lime-scale. Cleaning should be carried out to remove any impurities that could cause a lack of
pressure in the sterilisation chamber and possible cuts in the seal.
ATTENTION: do not allow residues of lime-scale or dirt to accumulate on the seal, since these
can damage or break it over time.
To maintain the unit in good working order, periodically clean
all the external parts using a soft cloth and normal neutral
detergents or just water (do not use abrasive products).
DO NOT use solvents that could damage the external plastic components of the unit.
DO NOT wash the unit with direct sprays or high-pressure jets or water, since any
infiltration into the electrical components could prejudice the working of the machine and the
safety systems.
Emptying and cleaning the tanks
CAUTION: DISCONNECT POWER SUPPLY. Non-observance may cause serious injury to
people or may seriously damage the unit.
ATTENTION: if the unit is not used for more than three days, both tanks should be emptied to
prevent deposits from forming.
Fig. 9.2-3