E8_Aus_rev07 2021-02-02
49
1. Empty the clean water tank: fit the end of the tube with the connector into the connector at the bottom of
the front panel (1 of Fig. 9.2-4), and the other end into an empty container.
2. Empty the internal tank for collecting used water: fit the end of the transparent tube into the connector at
the bottom of the front of the unit (2 of Fig. 9.2-4) and the other end into an empty container.
3. At the end of the draining operation, remove the tube from the connector by pressing its button.
Fig. 9.2-4
4. Remove the cover in order to access the tanks:
- lift the cover by 45° (Fig. 9.2-5) and pull it towards you (Fig. 9.2-6).
Fig. 9.2-5
Fig. 9.2-6
5. Carefully clean the tanks with the sponge supplied and water. Use the soft side of the sponge, not the
abrasive side. Clean with care, paying particular attention to any dirt that may have deposited in the
corners.
6. Remove the filters of the clean and dirty water tank (Fig. 9.2-7), rinse away any deposits under running
water and then reinstall them in the tank, taking care to position them correctly.