The table below lists all the alarm messages with the probable causes of faults; if your steriliser shows any of
the following error codes, before contacting the technical support service perform the checks indicated in the
table.
Anomalous change in
input voltage.
Fault in the mains
power supply or
inadequate input
socket.
Make sure the unit is connected to a
suitable mains power supply.
1. Temporary black-
out.
2. Two-pole thermal
safety switch
tripped.
3. Safety thermostat
tripped.
1. Wait for mains voltage to return.
2. Switch the unit back on. If the problem
persists contact the assistance service.
3. Allow the steriliser to cool for a few
hours and then reset the safety
thermostat at the front of the machine.
If the problem persists contact the
assistance service.
Excessive pressure
during sterilisation in
emergency cycles.
Fault during the
sterilization phase.
Allow the steriliser to cool and then perform
a sterilisation cycle with a small load in the
boiler (just one tray). If the problem persists
contact the assistance service.
Excessive pressure
during sterilisation.
Fault during the
sterilization phase.
Allow the steriliser to cool and then perform
a sterilisation cycle with a small load in the
boiler (just one tray). If the problem persists
contact the assistance service.
Insufficient pressure
during sterilisation.
Sterilisation phase
fault, or overload, or
steam loss.
Allow the steriliser to cool and then perform
a sterilisation cycle with a small load in the
boiler (just one tray). Make sure no water
leaks or drips on the front. If the problem
persists contact the assistance service.
Excessive
temperature during
sterilisation.
Fault during the
sterilization phase.
Allow the steriliser to cool and then perform
a sterilisation cycle with a small load in the
boiler (just one tray). If the problem persists
contact the assistance service.
Insufficient
temperature during
sterilisation.
Fault during the
sterilisation phase,
probably due to a leak
during a vacuum
phase.
Allow the steriliser to cool and then perform
a vacuum test. If this is successful, perform
a sterilisation cycle with a small load in the
boiler (just one tray). If the problem persists
contact the assistance service.
Unsaturated steam
during sterilisation.
Fault during the
sterilisation phase,
probably due to a leak
during a vacuum
phase.
Allow the steriliser to cool and then perform
a vacuum test. If this is successful, perform
a sterilisation cycle with a small load in the
boiler (just one tray). If the problem persists
contact the assistance service.
Cannot reach the
cycle vacuum
threshold.
Hydraulic leak during
the vacuum phase or
overloading.
Allow the steriliser to cool and then perform
a vacuum test. If the problem persists
contact the assistance service.
Cannot reach the
cycle pressure
threshold.
Fault in the steam
generation system or
hydraulic leak or
overloading.
Perform a sterilisation cycle with a small
load in the boiler (just one tray). If the
problem persists contact the assistance
service.