E8_Aus_rev07 – 2021-02-02
Fault in the door locked
position switch.
Perform a sterilisation cycle. If the problem
persists contact the assistance service.
Steam generator
inactive.
Fault in the steam
generator or electronic
board or steam
generator safety
thermostat tripped.
Contact the assistance service.
Temperature of steam
generator too high.
Fault in the electronic
board or temperature
sensor.
Switch off the steriliser and allow it to cool
for a few hours, then perform a sterilisation
system. If the problem persists contact the
assistance service.
Temperature of upper
band too high.
Fault in the electronic
board or temperature
sensor.
Switch off the steriliser and allow it to cool
for a few hours, then perform a sterilisation
system. If the problem persists contact the
assistance service.
Temperature of lower
band too high.
Fault in the electronic
board or temperature
sensor.
Switch off the steriliser and allow it to cool
for a few hours, then perform a sterilisation
system. If the problem persists contact the
assistance service.
Temperature of lower
band heater too high
in the pressure
maintenance cycle.
Fault in the electronic
board or temperature
sensor.
Switch off the steriliser and allow it to cool
for a few hours, then perform a sterilisation
system. If the problem persists contact the
assistance service.
Condensation battery
temperature too high
Sensor, connection, or
fan fault.
Contact the assistance service.
Problems about write
on Sd card
SD card inserted after
power on or removed
during a cycle
Switch off the autoclave and check if the sd
card is in the right place. If the problem
persists contact the assistance service
Water injections
finished.
Boiler overloaded or
clogged with scale or
water injection pump
inefficient.
Perform a sterilisation cycle with a small
load in the boiler (just one tray). If the
problem persists contact the assistance
service.
Water not being
supplied by the
Aquafilter® deioniser.
Error in the hydraulic or
electrical connections
with Aquafilter or
electronic fault.
Make sure that the Aquafilter connections
are correct and that there are no crushed or
bent pipes. Make sure the Aquafilter
®
inlet
tap is open. If the problem persists contact
the assistance service.
Problem in
transferring data
from power board
to display
Turn the autoclave off. Then, turn it on. If
the problem persists, contact technical
assistance.
Problem in
transferring data
from display to
power board
Turn the autoclave off. Then, turn it on. If
the problem persists, contact technical
assistance.