E8_Aus_rev07 – 2021-02-02
The following table shows the warning messages given by the steriliser using symbols or codes when it
detects a problem that prevents a cycle from starting.
Clean water under minimum
level.
Fill the clean water tank with distilled or
deionized water.
Used water at maximum
level.
Empty the used water tank.
The conductivity read by the
Aquafilter
®
deionizer is out of
range and the automatic
supply of water is therefore
impossible.
Replace the cartridges in the Aquafilter
deionizer.
An attempt was made to start
a cycle with the door open.
Before running a cycle, close the door.
The steriliser temperature is too high for the
requested operation. Switch it off and leave it
to cool leaving the door open.
The steriliser is not detecting
the SD memory card, or the
SD card is write-protected.
Check that the SD memory card is present
and correctly installed. Switch off the steriliser,
remove the SD memory card and check that
the safety switch allows writing.
The conductivity read by the
conductivity meter on the
autoclave is at the limits of
the acceptable values.
Empty the clean water tank as soon as
possible, and refill it with demineralised or
distilled water of the best quality.