14
168
Eversense E3 CGM User Guide
Alerts and Notifications Displayed on the Apple Watch (continued)
Apple Watch
Display
Description
Smart Transmitter Error
Appears when the system’s internal checks detect
a smart transmitter error.
No glucose readings are displayed until the error is
corrected.
Sensor Check
Appears once when the system’s internal checks
detect instability with the sensor which requires
a return to calibration Initialisation Phase.
Actions
Confirm the alert on your Eversense CGM
Mobile App. Follow the steps shown in the
Troubleshooting section to reset your smart
transmitter. If you are unable to complete the
reset, contact Customer Support.
Confirm the alert on your Eversense CGM Mobile
App. In the Initialisation Phase, you must perform
4 fingerstick calibrations spaced 2-12 hours apart.
Display of glucose readings will resume after the
2nd successful fingerstick calibration.
Alerts