18
192
Eversense E3 CGM User Guide
Q: I just received a new smart transmitter. How do
I unpair the old one and connect the new one to
my system?
A: On the Main Menu, tap Connect. Tap and hold
the name of your old smart transmitter. Tap OK to
stop the app from automatically connecting with
the old smart transmitter. Forget the old smart
transmitter from your phone's Bluetooth settings.
Follow the steps in this User Guide for pairing the
new smart transmitter with the app and linking it
to your sensor.
Calibration
Q: What time should I enter on the CALIBRATE
screen when I am notified to calibrate?
A: Enter the time you tested your blood glucose with
your meter. You must enter the blood glucose
reading within 10 minutes of doing the test.
Q: Why was my calibration rejected?
A: The system will reject the calibration for the any of
the following reasons:
• The blood glucose reading entered is less than
40 mg/dL.
• The blood glucose reading entered is greater than
400 mg/dL.
If another calibration is needed, the system will
prompt you.