19
213
Eversense E3 CGM User Guide
213
Eversense E3 CGM User Guide
Table 8 – Concurrence with YSI Range All Days in the PROMISE Study
Percent of Matched Pairs in Each YSI Glucose Range for
Each CGM Glucose Range YSI (mg/dL)
CGM
(mg/dL)
Number
of Paired
CGM-YSI
< 40 40-60 61-80 81-120 121-160 161-200 201-250 251-300 301-350 351-400 > 400
40 - 60 574 0% 62% 36% 1% 0% 0% 0% 0% 0% 0% 0%
61 - 80 1,178 0% 19% 66% 15% 0% 0% 0% 0% 0% 0% 0%
81 - 120 2,066 0% 0% 11% 76% 12% 0% 0% 0% 0% 0% 0%
121 - 160 2,067 0% 0% 0% 14% 71% 14% 0% 0% 0% 0% 0%
161 - 200 1,648 0% 0% 0% 0% 21% 66% 13% 0% 0% 0% 0%
201 - 250 1,437 0% 0% 0% 0% 1% 15% 68% 15% 1% 0% 0%
251 - 300 1,353 0% 0% 0% 0% 0% 1% 17% 56% 25% 1% 0%
301 - 350 1,191 0% 0% 0% 0% 0% 0% 1% 17% 70% 12% 0%
351 - 400 520 0% 0% 0% 0% 0% 0% 0% 3% 51% 42% 5%
> 400 155 0% 0% 0% 0% 0% 0% 0% 1% 21% 52% 26%
Eversense E3 Concurrence with YSI Values
The shaded area in the tables below show what percentage of YSI reference values were in the same range as the
sensor glucose values. As an example, when sensor glucose is between 81 and 120 mg/dL, YSI reference values are
in the same range 76% of the time.