55
Eversense E3 CGM User Guide
6
6. The MY GLUCOSE screen
appears with a grey blood drop
icon to identify your fingerstick
calibration is in progress.
The blood drop will turn red
when the calibration value
is accepted.
IMPORTANT: The smart transmitter should not be removed from over the sensor for at least 5 minutes
before the test to 15 minutes after the test while calibration is in progress. The Status Bar at the top of the screen
lets you know when calibration will be complete. If the smart transmitter is removed before the calibration is
complete, you will be prompted to calibrate again.
4. The CALIBRATE screen
now shows the time and
glucose reading you entered.
If not correct, repeat steps 3.
When correct, tap Submit.
5. The CALIBRATION
IN PROGRESS screen
appears.
Tap OK.

Table of Contents