How to remove the protection braid
1.
Release the angle lock lever of the scope and straighten the curved part of the
scope.
2.
Loosen the clamping screw of the braid stopper and slowly pull out the
protection braid from the scope.
CAUTION:
l
Do not pull out the protection braid with the scope angle lever locked or with your finger
on the angle lever.
l
The scope may be damaged. In order to protect the scope insertion part, it is
recommended to keep the protection braid attached to the scope insertion part at all
times.
3.8 Pre-operation/post-operation inspection
NOTE:
l
Check the inspection items even after using the machine.
Inspection Item Troubleshooting
whole
Is there foreign matter such as dust or sand
adhering to the insertion tube or each unit?
"Maintenance " (page
45)
Are there any cracks or scratches on the
insertion tube or each unit?
Replacement is
required.
Please contact your
dealer, our branch or
sales office.
Are there any cracks, deformations, or other
abnormalities in the exterior materials or
buttons?
Are there any abnormalities such as cuts or
buckling in the cable?
Carrying
case
Is there any damage or looseness on the
exterior parts of the carrying case, handles,
latches, etc.?
Please contact your
dealer, our branch or
sales office.
Are you using a dedicated carrying case?
If you use a case other than the dedicated
case, the product may be damaged or
malfunction.