7.1 Operating environment
Item Specifications
Operating temperature range
Inserting part of flexible
scope unit
In air: -25 to 80°C
Underwater: 10 to 30°C
Insertion part of rigid scope
unit
Parts other than the above
In air: -10 to 40°C
Working pressure
All parts In air: normal pressure (1,013 hPa)
Operating environment, altitude
All parts Altitude up to 2,000m
Operating environment, humidity
All parts 15 to 90% (relative humidity)
Liquid resistance
All parts
Even if machine oil, light oil, or 5% salt water adheres to the
product, functionality will not be affected.
Waterproof
All parts Tip: Equivalent to IPX7
Control unit: equivalent to IP52
Protection box: Dust-proof and drip-proof structure
equivalent to IP52
NOTE:
l
The operating environmental performance of this device does not guarantee it will
remain free of damage or failure.
7. Specifications