Safety Information
MaxTester Series 15
Electrical Safety Information
Maximum operation altitude 2000 m (6562 ft) (unit connected to external power
supply)
5000 m (16405 ft) (unit operated from battery)
Pollution degree
2 (unit connected to external power supply)
3 (unit operated from battery)
d
Overvoltage category I
Input power
e
unit: 8-16V;2A
AC adapter: 100 - 240 V; 50/60 Hz; 0.7 A
a. The battery will not recharge when the ambient temperature is below 5 °C (41 °F) or above 40 °C (104 °F).
Depending on the ambient temperature, at maximum power, the battery recharge may be interrupted
automatically after 15 minutes.
Operating time is dependent upon power consumption and temperature. If the unit is used at maximum
power and maximum rated temperature, to prevent any damage to the battery, the unit will be powered
off automatically after 15 minutes.
b. To optimize the performance of the battery, you should store the unit in a cool dry place.
c. Measured in 0 °C to 31 °C (32 °F to 87.8 °F) range, decreasing linearly to 50 % at 40 °C (104 °F).
d. Equipment must be normally protected against exposure to direct sunlight, precipitation and full wind
pressure.
e. Not exceeding ± 10 % of the nominal voltage.
CAUTION
The use of voltages higher than those indicated on the label affixed
to your unit may damage the unit.
CAUTION
The operation and storage temperatures, as well as the altitude and
relative humidity values of some modules may differ from those
specified for your platform. In this case, always ensure that you
comply with the most restrictive conditions (either module or
platform).
Equipment Ratings