EasyManua.ls Logo

Fluke 196 - Page 101

Fluke 196
116 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications
Automatic Scope Measurements
9
93
Power
Power Factor ..................... ratio between Watts and VA
Range......................................................0.00 to 1.00
Watt................................. RMS reading of multiplication
corresponding samples of input A (volts)
and Input B (amperes)
Full Scale Reading .................................... 999 counts
VA ............................................................ Vrms x Arms
Full Scale Reading .................................... 999 counts
VA Reactive................................................
((VA)
2
-W
2
)
Full Scale Reading .................................... 999 counts
Phase
Range..........................................-180 to +180 degrees
Resolution....................................................... 1 degree
Accuracy
0.1 Hz to 1 MHz ........................................
±1 degrees
1 MHz to 10 MHz ......................................
±3 degrees
Temperature (TEMP)
With Optional Temperature Probe
Ranges (
°C or °F) ............................... -40.0 to +100.0 °
-100 to +250 °
-100 to +500 °
-100 to +1000 °
-100 to + 2500 °
Probe Sensitivity ..........................1 mV/°C and 1 mV/°F
Decibel (dB)
dBV..............................................dB relative to one volt
dBm................. dB relative to one mW in 50
or 600
dB on........................................VDC, VAC, or VAC+DC
Accuracy.........................same as VDC, VAC, VAC+DC

Table of Contents

Related product manuals