Loading...
Fluke 192/196/199
Users Manual
88
Dual Input Oscilloscope
Isolated Inputs A and B (Vertical)
Bandwidth, DC Coupled
FLUKE 199....................................... 200 MHz (-3 dB)
FLUKE 196....................................... 100 MHz (-3 dB)
FLUKE 192......................................... 60 MHz (-3 dB)
Lower Frequency Limit, AC Coupled
with 10:1 probe ......................................<2 Hz (-3 dB)
direct (1:1) .............................................<5 Hz (-3 dB)
Rise Time
FLUKE 199....................................................... 1.7 ns
FLUKE 196....................................................... 3.5 ns
FLUKE 192....................................................... 5.8 ns
Analog Bandwidth Limiters.............. 20 MHz and 10 kHz
Input Coupling ................................................... AC, DC
Polarity................................................ Normal, Inverted
Sensitivity Ranges
with 10:1 probe ............................50 mV to 1000 V/div
direct (1:1) ...................................... 5 mV to 100 V/div
Trace Positioning Range..............................
±4 divisions
Input Impedance on BNC
DC Coupled .....................1 M
(±1 %)//15 pF (±2 pF)
Max. Input Voltage
with 10:1 probe ......................................600 V CAT III
1000 V CAT II
direct (1:1) .............................................300 V CAT III
(For detailed specifications, see Safety”)
Vertical Accuracy....................
±(1.5 % + 0.04 range/div)
Digitizer Resolution................... 8 bits, separate digitizer
for each input
Horizontal
Maximum Time Base Speed:
FLUKE 199 .....................................................5 ns/div
FLUKE 196 .....................................................5 ns/div
FLUKE 192 ...................................................10 ns/div
Minimum Time Base Speed (Scope Record) ....2 min/div
Real Time Sampling Rate (for both inputs simultaneously)
FLUKE199:
5 ns to 2
µs /div................................. up to 2.5 GS/s
5
µs to 120 s/div..........................................20 MS/s
FLUKE 196:
5 ns to 2
µs /div.................................... up to 1 GS/s
5
µs to 120 s/div..........................................20 MS/s
FLUKE 192
10 ns to 2
µs /div.............................. up to 500 MS/s
5
µs to 120 s/div..........................................20 MS/s