37
Table 21. Voltage Function Verification: DC Voltage into a 1 MΩ Load (cont.)
Nominal Value (dc) Measured Value (dc) Deviation (mV) 1-Year Spec. (mV)
-450.0 mV
1.22
500.0 mV
1.35
-500.0 mV
1.35
3.3 V
8.35
-3.3 V
8.35
4.0 V
10.10
-4.0 V
10.10
33.0 V
82.60
-33.0 V
82.60
Table 22. Edge Function Verification
Nominal Value (p-p) Frequency Pulse Response Time (nS) 1-Year Spec. (ps)
25.0 mV 1 MHz
400
250.0 mV 1 MHz
400
250.0 mV 10 kHz
400
250.0 mV 100 kHz
400
250.0 mV 1 MHz
400
2.5 V 1 MHz
400
Table 23. Wave Generator Function Verification: 1 MΩ Load
Waveform
Nominal
Value (p-p)
Frequency
Measured
Value (p-p)
Deviation
(mV)
1-Year Spec.
Square 5.0 mV 10 kHz 0.25 mV
Square 20.0 mV 10 kHz 0.70 mV
Square 89.0 mV 10 kHz 2.77 mV
Square 219.0 mV 10 kHz 6.67 mV
Square 890.0 mV 10 kHz 26.80 mV
Square 6.5 V 10 kHz 195.10 mV
Square 55.0 V 10 kHz 1.65 V
Sine 5.0 mV 10 kHz 0.25 mV
Sine 20.0 mV 10 kHz 0.70 mV
Sine 89.0 mV 10 kHz 2.77 mV
Sine 219.0 mV 10 kHz 6.67 mV
Sine 890.0 mV 10 kHz 26.80 mV