5080A
Service Manual
36
Table 20. Voltage Function Verification: DC Voltage into a 50 Ω Load (cont.)
Nominal Value (dc) Measured Value (dc) Deviation (mV) 1-Year Spec. (mV)
-25.0 mV
0.16
55.0 mV
0.24
-55.0 mV
0.24
100.0 mV
0.35
-100.0 mV
0.35
220.0 mV
0.65
-220.0 mV
0.65
250.0 mV
0.72
-250.0 mV
0.72
550.0 mV
1.47
-550.0 mV
1.47
700.0 mV
1.85
-700.0 mV
1.85
2.2 V
5.60
-2.2 V
5.60
Table 21. Voltage Function Verification: DC Voltage into a 1 MΩ Load
Nominal Value (dc) Measured Value (dc) Deviation (mV) 1-Year Spec. (mV)
0.0 mV
0.10
5.0 mV
0.11
-5.0 mV
0.11
22.0 mV
0.15
-22.0 mV
0.15
25.0 mV
0.16
-25.0 mV
0.16
45.0 mV
0.21
-45.0 mV
0.21
50.0 mV
0.23
-50.0 mV
0.23
220.0 mV
0.65
-220.0 mV
0.65
250.0 mV
0.72
-250.0 mV
0.72
450.0 mV
1.22