729/729FC
Users Manual
16
The screen has two sections:
The upper display shows the internal pressure controller
status and includes:
Mode: Measure, Source, and Vent
Pressure Value: Current measured pressure and unit
Status: Shows nothing in measure mode, NOT READY,
STABLE in source mode, and VENTING, VENTED in
vent mode.
Setpoint: Current setpoint for source function.
Step Size: The step size when (Step Up) and
(Step Down) are pushed when sourcing
pressure.
The lower display shows the electrical status and includes:
Mode: Measure Current, Source Current, Simulate
Current, Measure V dc, and Switch.
Reading: Current reading from measure functions or
setpoint for source functions.
Switch State: Shows the pressure reading when the
switch contacts open and close.
Percent: Shows the percentage value of 4 mA to 20 mA
or 1 V to 5 V in measure functions.
Status: Shows Open circuit detected in mA source
mode.

Table of Contents