EasyManua.ls Logo

Fluke 729 - Pressure Transmitter (P;V); Pressure Transmitter (P;P)

Fluke 729
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
729/729FC
Users Manual
40
Pressure Transmitter (P/V)
Use this feature to calibrate a pressure to voltage
transmitter with voltage measurement. For connections, see
Figure 11.
1. From page 1 of the Tasks menu, push the arrows to
highlight Pressure Transmitter (P/V).
2. Push or .
3. From this screen, use the arrows, numeric keypad, and
to change the values of the source pressure and
measure voltage.
Push (Continue) to edit these other
parameters:
Tolerance of Range
Test Strategy
Auto Settling Time
Transfer Function (Linear or Square
Root).
Push (Save as Customized) to save these
settings as a customized task. Recall these customized
tasks from the Tasks menu at any time.
Pressure Transmitter (P/P)
To calibrate a pressure-to-pressure transmitter, use this
feature. This feature requires a pressure module to
measure pressure output from a connected transmitter.
To use the function, see Figure 15:
1. From page 1 of the Tasks menu, push the arrows
to highlight Pressure Transmitter (P/P).
2. Push or .
3. From this screen, use the arrows, numeric keypad,
and to change the values of the source
pressure and measure pressure.
4. Push (Continue) and these parameters
can be edited:
Tolerance of Range
Test Strategy
Auto Settling Time
5. Push (Save as Customized) to save these
settings as a customized task. Recall this
customized task from the Tasks menu at any time.