Automatic Pressure Calibrator
Maintenance
67
Error Messages
The Product display shows error messages when the
Product fails to meet certain conditions or there is a
problem with the configuration of the Product or test. See
Table 8 to troubleshoot these issue.
Table 8. Error Messages
Error Message Explanation Potential Root Cause and Solution
Too many errors Too many errors, the error queue overflow.
Clear the errors with the Product user
interface
Cannot read file Failed to read file from file system.
1. File corrupted, delete the file and try
again.
2. Data flash memory failure, the main
PCA needs service.
Cannot recall procedure Failed to recall downloaded procedure.
1. The downloaded procedure contains an
illegal configuration.
2. File corrupted, download the procedure
again.
Cannot store results Failed to store task results.
1. Not enough space, delete some files
and run the task again.
2. Delete all results and run the task again.