Automatic Pressure Calibrator
Measure Mode
49
Trim to Applied Values
1. Enter the LRV pressure value necessary in the Setpoint
entry and let the pressure settle.
2. Push (Trim LRV) to trim to the lower range of
pressure applied.
3. Enter the URV pressure value necessary in the
Setpoint entry and let the pressure settle.
4. Push (Trim URV) to trim to the upper range of
pressure applied.
As Left
Push (As Left). Repeat the procedure from As
Found.
Calibration results save to the Test Results menu.
Find Task by Tag
Downloaded tasks from DPCTrack2 or other supported
software are shown. Use the up and down arrows to scroll
through the tasks. Push (Page Up) or
(Page Down) where necessary. Push (Recall) to
recall the highlighted task.
Bus Polling
Bus Polling goes back into the bus polling screen and scans
for connected HART devices.
If no HART device is found after Polling has completed,
push (Restart Bus Polling) to restart bus polling to
search of the Product BUS for HART devices. The Product
locates the device.
Measure Mode
The measure mode has these functions:
mA dc measure
mA dc source
mA dc simulate
V dc measure
Switch (continuity)
The lower left of the display shows the operation mode.