Automatic Pressure Calibrator
Setup Menu
29
Manage Custom Tasks
Use this submenu to delete custom tasks (see Tasks
Menu). To delete a single task, use the arrow to select a
custom task and push (Delete). To delete all of
the tasks, push (Delete All).
Maintenance Menu
Maintenance has these sections:
• Drain Water (condensation)
• Keypad Test
• Pressure Sensor Characterize (see Calibration
Manual)
• Supply Sensor Self Calibration (see Calibration
Manual)
• Exhaust
Drain Water (Condensation)
To drain condensation from the Product, see Figure 10:
1. Connect the hose to collect drained water.
2. Push (Continue).
3. The display shows Creating pressure, please
standby. The pump works to establish a pressure.
4. To let the water out, use a small-tipped tool, such as
a small Philips screwdriver, to push the DRAIN
button on the side of the Product.
5. Push to repeat the drain operation.
6. When finished with water drain, push in on the fitting
to release the hose.