Automatic Pressure Calibrator
Tasks Menu
39
Pressure Leak Test
Use this feature to test a pressure device for leaks. For
connections, see Figure 15:
1. From this screen, use the arrows, numeric keypad, and
to change the values of these parameters:
• Setpoint (Pressure value)
• Settling Time
• Test Time
Note
Make sure to push after EACH value change.
2. Push (Continue). When the pressure leak test
completes, the Product shows the results. Results
include Start Pressure, End pressure, Test time, and
Leak rate.