729/729FC
Users Manual
36
Current to Pressure Test (I/P)
Use this feature to calibrate a current to pressure (I/P)
converter with current source and pressure
measurements. See Figures 13 and 14:
1. From page 1 of the Tasks menu, push the arrows to
highlight Current to Pressure Test (I/P).
2. Push or .
3. From this screen, use the arrows, numeric keypad,
and to change the values of the source current
and measure pressure.
Note
Make sure to push after EACH value
change.
4. Once you enter the values, push (Continue).
5. Tolerance of Range, Test Strategy, and Auto Settling
Time can be set from here.
To save these settings as a customized task, push
(Save as Customized). You can then recall
the saved task from the Tasks menu.
6. Once you enter the parameters, push
(Continue).
7. Push (Manual Test) or (Auto Test).
8. Adjust with to go to 0%, 50%, and 100%.
9. Push (As Left) to do the As Left test.