729/729FC
Users Manual
34
Pressure Switch (SW)
Use this feature to calibrate a pressure switch. See
Figure 12 for connections:
1. From page 1 of the Tasks menu, push the arrows to
highlight Pressure Switch (SW).
2. Push or .
3. From this screen, use the arrows, numeric keypad,
and to change the values of these parameters:
Setpoint
Tolerance
Min Deadband
Max Deadband
Max Pressure
Switch Type
Note
Make sure to push after EACH value
change.
You can save these settings as a customized task.
Push (Save as Customized). Recall this
customized task from the Tasks menu.
4. Once you enter the parameters, push
(Continue).
5. Push (Auto Test) or (Manual Test)
to do the As Found switch test. The test runs.
After the test runs, the results show on the display.
To save the test results:
1. Push (Done).
2. Push to edit the Tag information.
3. Push (Done).
4. Move the cursor to the S/N field (Serial Number).
5. Push to edit the S/N information.
6. Push (Done).
7. Move the cursor to the User ID field.
8. Push (User List) to select a user.
9. Push (Done).
10. Adjust the switch and push (Step Up)
and (Step Down) to verify.
11. Push (As Left) to do the As Left test.