729/729FC
Users Manual
46
HART Service
HART Service includes these submenus:
• PV Zero Trim
• mA Output Trim
• Trim to Applied Values
• Set Fixed mA Output
• Re-range Transmitter
• Device Diagnostics
Highlight the HART Service menu selection and push
(Continue). When you enter some of the
submenus, a warning tells you to remove the loop from
AUTOMATIC control. Push (Continue).
PV Zero Trim
This function sets the digital pressure value of the
transmitter to zero.
1. Push (Trim Zero) to trim the PV output value to
zero.
2. Push (Done) when completed.
mA Output Trim
In the mA output trim screen, you can adjust the transmitter
output.
1. Start with the 4 mA range. Push (Fetch) to get
the current measured mA value.
2. Push (Trim 4 mA) to send the measured value
for correction. Push (Next Point). The Product
moves to the 20 mA range.
3. Repeat steps 1 and 2.
4. When complete, push to return to the previous
point or (Done).
Trim to Applied Values
1. Enter the LRV pressure value necessary in the Setpoint
entry and then push and let the pressure settle.
2. Push (Trim LRV) to trim the lower range of
pressure applied.
3. Enter the necessary URV pressure value in the
Setpoint entry and push and let the pressure
settle.
4. Push (Trim URV) to trim the upper range of
pressure applied.