Manage Test Results
The Product saves test results as found and as left and
saves calibration results.
1. In the Setup menu, push the arrow buttons to
highlight Manage Test Results.
2. Push to show a list of tests. The tests are
designated by date and time, Tag ID, and serial
number.
3. Push the arrow buttons and or to view the
selected test results.
4. Push to select the As Found or As Left test
result.
5. Push to redo the calibration and save a new
test result.
To delete all test results:
1. Go to the Manage Test Results screen.
2. Push (Delete All).
3. Confirm to delete all records.
4. Push (OK) to delete all records or
(Cancel) to escape the deletion process.
Manage Screen Shots
The Product can take screen shots of any screen and
store them.
To take a screen shot:
1. Bring up the screen to save.
2. Push and hold . The Product shows File Saved.
To view the screen shots:
1. From the Setup menu, push the arrows to highlight
Manage Screen Shots.
2. Push or . The Product shows a list of
screenshots and their time and date stamps.
3. Use the arrows to highlight a screen shot.
4. Push or and the screen shot is shown. The
screen shows: “You are reviewing a screen copy!”
over the saved screen.
5. Push to delete the individual screen.

Table of Contents