Users Manual
42
HART Functionality
The Product interfaces with HART devices. It can:
• View and modify transmitter setup and data that
includes:
o Write Lower Range Value (LRV)
o Write Upper Range Value (URV)
o Write PV Unit
o Write Message and Descriptor
o Write Damping Value
o Write Date
• Perform service features including:
o PV Zero Trim
o mA Output Trim
o Trim to Applied Values
o Set Fixed mA Output
o Re-range Transmitter
o Device Diagnostics
• Perform HART calibration without using a Task (Ad
hoc)
• Perform HART calibration from a downloaded Task