Automatic Pressure Calibrator
Setup Menu
25
729 Information
The 729 Information submenu is an information-only
screen that shows information about the Product. The
screen shows:
• The Product model (729 or 729 FC)
• Serial Number
• Firmware Revision
• The last calibration date of the Product pressure
components.
• The last calibration date of the Product electrical
components.
There are also softkeys at the bottom of this screen.
These are:
• - Calibrate Pressure – Follow the prompts
on the screen and the Product takes you through a
procedure that calibrates the pressure components
of the Product. Enter the Product password to use
this function. (Options) lets you save the
calibration, go back a step, skip a step, or move to
the next section. See Calibration Manual for details.
• - Calibrate Electrical – Follow the prompts
on the screen and the Product takes you through a
procedure that calibrates the electrical components
of the Product. Enter the Product password to use
this function. (Options) lets you save the
calibration, go back a step, skip a step, or move to
the next section. See Calibration Manual for details.
• - Battery Information – An information-only
screen that shows if the battery is present, the
remaining battery charge level, battery temperature,
and battery voltage.
• - Change Password – Use to change the
password for advanced functions.